For:      Adv. No. 19-064        –          SUPPLY, DELIVERY AND INSTALLATION OF ONE (1) UNIT X-

RAY MACHINE FOR USE IN TUBURAN DISTRICT HOSPITAL.

 

This addendum is issued to modify or amend items in the Bid Documents for the procurement of the different projects of the Province of Cebu, as listed.  This shall form an integral part of the Bid Documents.

 

  • TECHNICAL SPECIFICATIONS:

 

X-RAY MACHINE, brand new

 

40 kW / 125 kVp Hi -Frequency Line Power Generator, Single Phase line power

• Digital Imaging Ready, ULTRA High Frequency Power, 120 kHz PLUS

• 50 kW maximum output; (according to IEC 60601)

• mA Range: 10 to 500

• kVp Range: 40 to 125 kVp, in 1 kVp step

• mAs Range: 0.1 – 500mAs

• Timer Range: 0.001 – 10 seconds, 38steps

• User programmable max 1,280 programs with APR utility software

• Technique selection; 4 point display (kV, mA, Time, mAs)

• Self-Diagnostics, Anode Heat Monitor, Error Messaging, Auto Shut-off Timer, History Reporting Log, RS-232 Port

• STORED ENERGY technology permits full operation using a standard wall outlet or alternate power source

* Input power (50/60 Hz): 230V~

Reproducibility Coefficient of Variation: kV < 0.005, Time < 0.005, mAs < 0.01

Accuracy kV < ±(1%+1kV), mA < ±(3%+1mA), Time < ±(1%+0.5ms), mAs < ±(3%+0.1mAs)

Linearity Coefficient of Linearity < 0.01 : CL = (X1-X2)/(X1+X2), where X is mR/mAs

Dimension 622 x 652 x 405mm (W x H x D)

Weight 100kg

X-Ray Tube:  Radiographic X-Ray Tube

• 0.6/1.5 mm focal spot sizes with 200,000 Heat Unit capacity

• 125 kVp, Standard speed rotor, 14° anode target angle, 90° arms

*More than 2.9mmAl eq. at 80kVp

High Voltage Cables: One Pair, 20 ft long (6 meters) with Federal Terminals.

(Tube Stand Floor Mounted)

Tube Rotation Angle: ±135°

Tube stroke; Longitudinal=2200mm; Lateral=220m; 1,580mm (440~2,020mm from floor to focus)

Lock(Brake); EM Lock, Switch on/off

Balance: Counterweight

Column Rotation 90° step

Electrical Rating 24VDC, 3A

Dimension / Weight 2,317(H) x 1,373(D) mm x 3,006(W) / 240kg (529lbs)

PBT-4(Patient Bucky Table) 4- way floating table

Longitudinal 1,000(±500)mm

Tabletop

Transverse(Lateral) Movement 250(±125)mm

Bucky Longitudinal

350mm with standard tray

Inherent Filtration 1.2mmAl at 100kV

 

X-RAY MACHINE, brand new

 

40 kW / 125 kVp Hi -Frequency Line Power Generator, 3-Phase line power

• Digital Imaging Ready, ULTRA High Frequency Power, 120 kHz PLUS or Auto High Frequency Power

• 50 kW maximum output; (according to IEC 60601)

• mA Range: 10 to 630

• kVp Range: 40 to 150 kVp, in 1 kVp step

• mAs Range: 0.1 – 500mAs

• Timer Range: 0.001 – 10 seconds, 38steps

• User programmable min=270 (x20) programs with APR utility software

• Technique selection; 4 point display (kV, mA, Time, mAs)

• Self-Diagnostics, Anode Heat Monitor, Error Messaging, Auto Shut-off Timer, History Reporting Log, RS-232 Port

• STORED ENERGY technology permits full operation using a standard wall outlet or alternate power source

* Input power (50/60 Hz): 230V~

Reproducibility Coefficient of Variation: kV < 0.005, Time < 0.005, mAs < 0.01

Accuracy kV < ±(1%+1kV), mA < ±(3%+1mA), Time < ±(1%+0.5ms), mAs < ±(3%+0.1mAs)

Linearity Coefficient of Linearity < 0.01 : CL = (X1-X2)/(X1+X2), where X is mR/mAs

Dimension 622 x 652 x 405mm (W x H x D)

Weight 100kg

X-Ray Tube:  Radiographic X-Ray Tube

• 0.6/1.5 mm focal spot sizes with 200,000 (min.) Heat Unit capacity

• 125 (min.) kVp, Standard speed rotor, 14° anode target angle, 90° arms

*More than 2.9mmAl eq. at 80kVp

High Voltage Cables: One Pair, 20 ft long (6 meters) with Federal Terminals.

(Tube Stand Floor Mounted)

Tube Rotation Angle: ±135°

Tube stroke; Longitudinal=2200mm; Lateral=220m; 1,580mm (440~2,020mm from floor to focus)

Lock(Brake); EM Lock, Switch on/off

Balance: Counterweight

Column Rotation 90° step

Electrical Rating 24VDC, 3A

Dimension / Weight Manufacturer’s standard

PBT-4(Patient Bucky Table) 4- way floating table

Longitudinal 1,000(±500)mm (min.)

Tabletop

Transverse(Lateral) Movement 250(±125)mm

Bucky Longitudinal

350mm with standard tray

Inherent Filtration 1.2mmAl at 100kV

 

 

Tabletop Max. Patient Weight 300kg (660lbs)

Size 2,200(W) x 750(D) x 45(H) mm

Bucky Type; Oscillating Grid type

Standard tray for CR and Analog cassette

– Tray size: 17”x17” (for 8”x10” to 17”x17” cassette)

– Tray size: 14”x17” (only for 14”x17” cassette)

– Rotating: Center reference

Grid FD 100cm, 103 lpi, ratio 10:1

Lock(Brake) EM Lock, beam sensor on/off

Center indication Buzzer sound and LED

Electrical Rating 100–240Vac, 200VA, 50/60Hz

Dimension / Weight 2,200(W) x 750(D) x 660 mm / 150kg(330lbs)

Vertical Wall Stand(WBS)

Cassette stroke Vertical 1,640mm (420~2,060mm from floor to Bucky center)

Bucky Type; Grid type / Oscillating Grid type

Standard tray for CR and Analog cassette

– Tray size: 17”x17” (for 8”x10” to 17”x17” cassette)

Grid FD 150cm, 103 lpi, ratio 10:1

Inherent Filtration 0.5mmAl at 100kV

Lock(Brake) EM Lock, Switch on/off

Balance Counter Weight

Electrical Rating 24VDC, 1A

Dimension / Weight: Manufacturer`s Standard

Collimator

Model / Maker R108

Control Manual with 30sec. lamp timer

Field Shape Rectangular

Max. Field Size More than 43x43cm(17x17inch) at 100cm SID

Leakage Radiation Less than 100mR/hr

Max. kVp shield 150kV

Inherent Filtration Min. 2.0mmAl eq.

Luminosity Over 160LUX at 100cm SID (Typ. 250LUX)

Light source Single LED***

Standard Tape measure, Rotating flange

Preinstalled Option Laser line

Electrical Rating 20–30VAC, 30VA, 50/60Hz

Dimension / Weight: Manufacturer`s Standard

 

Accessories: Lead Gown, Gonadal Shield male & female, Thyroid Lead Shield, Patient Caliper

 

Warranty: One (1) Year

With Good Performance Certificate from any 2 Hospitals

Brand should be at least 10 years in the market.

With installation of the same unit in any hospital

 

*With demonstration and installation

*semi PMS

*Service Center w/ Cebu location or w/ agreement w/ 3rd party in Cebu

 

Tabletop Max. Patient Weight 300kg (660lbs)

Size:  Manufacturer’s standard

Bucky Type; Oscillating Grid type

Standard tray for CR and Analog cassette

– Tray size: 17”x17” (for 8”x10” to 17”x17” cassette)

– Tray size: 14”x17” (only for 14”x17” cassette)

– Rotating: Center reference

Grid FD 100cm, 103 lpi, ratio 10:1

Lock(Brake) EM Lock, beam sensor on/off

Center indication Buzzer sound and LED

Electrical Rating 100–240Vac, 200VA, 50/60Hz

Dimension / Weight 2,200(W) x 750(D) x 660 mm / 150kg(330lbs)

Vertical Wall Stand(WBS)

Cassette stroke Vertical 1,640mm (420~2,060mm from floor to Bucky center)

Bucky Type; Grid type / Oscillating Grid type

Standard tray for CR and Analog cassette

– Tray size: 17”x17” (for 8”x10” to 17”x17” cassette)

Grid FD 150cm, 103 lpi, ratio 10:1

Inherent Filtration 0.5mmAl at 100kV

Lock(Brake) EM Lock, Switch on/off

Balance Counter Weight

Electrical Rating 24VDC, 1A

Dimension / Weight: Manufacturer`s Standard

Collimator

Model / Maker R108

Control Manual with 30sec. lamp timer

Field Shape Rectangular

Max. Field Size More than 43x43cm(17x17inch) at 100cm SID

Leakage Radiation Less than 100mR/hr

Max. kVp shield 150kV

Inherent Filtration Min. 2.0mmAl eq.

Luminosity Over 160LUX at 100cm SID (Typ. 250LUX)

Light source Single LED***

Standard Tape measure, Rotating flange

Preinstalled Option Laser line

Electrical Rating 20–30VAC, 30VA, 50/60Hz

Dimension / Weight: Manufacturer`s Standard

Accessories: Lead Gown, Gonadal Shield male & female, Thyroid Lead Shield, Patient Caliper

*Compatible with existing portable DR X-RAY (MobileDiagnost Opta version 1.0, 125kv 110mA), with software

 

Warranty:  two (2) years

With Good Performance Certificate from any 2 Hospitals

Brand should be at least 10 years in the market.

With installation of the same unit in any hospital

 

*With demonstration and installation

*semi PMS

*Service Center w/ Cebu location or w/ agreement w/ 3rd party in Cebu

 

For guidance and information of all concerned.

 

Cebu Capitol, Cebu City, March 5, 2019.

 

(sgd)  ATTY. MARK C. TOLENTINO

Chairman

 

Conformed:

 

______________________________________

(Signature over printed name)

 

______________________________________

(Name of Company)

 

Date: _____________________________________

 

**Please acknowledge receipt hereof by re-sending this through Telefax No. (32) 253-9614, duly signed received.**

 

/2019/bac/addendum/11/specs/