1.

 

Cardiac Monitor with Defibrillator, with Table/ Cart

 

Technical Specifications: (minimum)

Modes

AED Manual, AED with manual override

Waveform Type

Rectilinear Biphasic Waveform (RBW)

Waveform Shape

Constant Current

Waveform Duration

Fixed – 10 msec

Joule Settings

1-360 Joules

VF – First Shock Efficacy

99% at 129 Joules

Charge time to max energy

Less than 7 seconds to 200 J with new fully charged battery

Impedance Compensation

Through the use of internal digital resistors

Synchronized Cardioversion Yes

AED to Manual Operation

Manual Interface available

Real CPR Help

CPR Quality Feedback System Rate and Depth – Metronome and Voice Prompts

MONITORING/PARAMETERS

Display Type: High Resolution EL or LCD

Screen Size: 14.37 cm

ECG Monitoring: 3 lead, 5 lead and 12 lead ECG cable

Standard Lead Choices: Pads/Paddles, I,II,III ( if 12 lead all leads)

12-Lead ECG: Yes

12-Lead ECG Algorhythm: Non Interpretive, simultaneous

ECG Size: 0.5, 1, 1.5, 2, 3 cm/mV

STEMI Management System: Yes

Capnography EtCO2: Respironic CapnoStat3 – Mainstream – Adults, Pediatrics and Neonates

NIBP: Motion Tolerant Technology

Alarms: Heart Rate & VF-VT

*ISO 13485; *Authorized or Exclusive Distributor;                *List of Installation;  *Biomedical Engineer with Training Certification from manufacturer

*Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 60 days; with demonstration

 

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

 

NO CHANGES

 

/2018/bid bulletin/#32/p2

 

2

 

ECG Machine with Analyzer

 

5.7 inch Touch Screen Operation and Patient Data Entry

Direct Connection with Printer

Mega Memory: Internal 200 + Removable USB 20,000/8G (Resting ECG Records)

Network Interface: LAN

Arrhythmia Monitoring & Summary up to 15 minutes

 

Recording Modes

Auto Mode

And you can freeze ECG on the display if any ECG abnormality is found, then you can start automatic recording and analysis of the captured ECG on the display.

Pre-Check Mode

Monitoring occurrence of arrhythmia. If arrhythmia is detected, ECG recording and analysis will be started automatically.

Manual Mode

Long Term Mode

Arrhythmia Analysis: Analyze ventricular premature contraction, supraventricular premature contraction and atrial fibrillation by recording selected single lead for a long term.

Autonomic Nervous System Test(ANST) can be performed.

-Brand must be in the local and international market for at least 10 years

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 60 days; with demonstration

 

Allocation:

CPH-Balamban = 1; CPH-Bogo =1; CPH-Danao = 1

 

NO CHANGES

 

3

 

Emergency Cart

 

Size: at least 470mm x 680mm x 980mm, 250mm

Cabinet height steel plate body

ABS Molded base and top place

Pull out writing table, push handle; Four drawers with manual and/or electronic lock

with removable dividers

with guard rubber bumper

Four (4) heavy duty caster wheels at least five (5) inches in diameter, locking brakes

With needle disposal box

2 ABS molded dust box

CPR Board

Warranty: 1 year warranty

Delivery: To be delivered directly to allocated hospital within 60 days

 

Allocation:

CPH-Balamban = 1;

CPH-Bogo = 1

CPH-Danao = 1

 

NO CHANGES

 

/2018/bid bulletin/#32/p3

 

4

 

3 Crank ICU Hospital Bed Adult with Bedside Table, IV Pole & Foot Stool

 

•Electric five functions inclinable hospital bed

•Sectional type elevating guardrails

•Powder coating mild steel frame

•Detachable ABS head and foot bed boards

•Central control locking castors

•Electric inclinable lifting system

•Size: 224 x 104 x 49

With leatherette mattress

Warranty: 1 year

Delivery: To be delivered and assembled directly to allocated hospital within 60 days

 

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

 

NO CHANGES

 

5

 

Infusion Pump with Stand

 

infusion modes including multi-step mode, delayed start mode and interval / intermittent mode which allows the appropriate infusion therapy for the patient.

 

Midpress system

Midpress peristaltic function

With Anti Free Flow (AFF) function, prevents inadvertent excess infusion by automatically closing the tube with the anti-free flow clip when the door is opened.

 

can detect any brand of IV sets

Volume Limit Range: 1-9, 999ml, in 1 mL increment

LED status indicators

Has alarm: sound and light when there is any unusualities during infusion

 

Brand must be in the local and international market for at least 10 years

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 60 days with demonstration

 

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

 

NO CHANGES

 

6

 

Fiber Optic Laryngoscope with Blades, heavy duty

 

With LED replacement lamps

Adult Blades: mckintosh size 3 & 4

Pedia Blades: miller (straight) size 0 & 1

Blades easily convert from lamp to fiber-optic illumination

Single-piece, stainless steel blade ensures integrity and minimizes corrosion Bright, white Halogen HPX illumination for clear, color-corrected view

Our free repolishing service makes these an affordable choice

Knurled finish ensure durability and a secure grip

Battery Operated

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 30 days

 

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

 

NO CHANGES

 

/2018/bid bulletin/#32/p4

 

7

 

Mechanical Ventilator with Air Compressor

 

Patient type: Pediatric, Adult

Control Principle:     Must be time-cycled, volume-controlled, pressure limited

Gas Supply: Must have Turbine technology

Turbine Exchange: Exchange interval must be 8 years, with no limit in operating hours during this interval

Flow sensor Type: Must be temperature-compensated hot-wire anemometer

Ventilation Modes:

Standard modes:     VC-CMV/VC-AC, VC-SIMV, SPN-CPAP

Optional modes:      VC-MMV, PC-APRV, PC-BIPAP/PC-SIMV+, PC-AC

Optional Enhancement:

• AutoFlow: Must have Automatic adaption of the inspiratory flow in volume orientated ventilation modes.

• NIV        Must be capable in providing Non Invasive Ventilation with optimized alarm systems and automatic leakage compensation.

• Capnography: Must be using Mainstream CO2 measurement

• MonitoringPlus      Must have Loops, Trends, user Logbook

• LPO      Must have Low Pressure Oxygen. Independent oxygen supply, e.g. with an O2 concentrator

• Nurse call  Must have Connection for transmitting alarm signals to a central, alarm system

Advanced Features:

• Oxygen enrichment – provides 180 seconds of 100% O2 prior to suction procedure

– interrupts ventilation and delivers a minimal flow to detect reconnection automatically

• Manual Inspiration/Inspiration Hold Provides manual ventilation

• Day/Night screen switch-over

Selection of two (2) types of screen for day and night mode

• Key pod lock prevents unwanted modification of settings

• Standby mode Interrupts ventilation and provides access to other ventilator options and settings

Alarms:    Must have an Audible and visual alarms with event log and priorities

Performance data:

Trigger:

Leak adapted Trigger: Flow/volume trigger

Setting Parameters:

Frequency: At least 5 to 20 Hz

Peak inspiratory pressure:      10 to 80 mbar

PEEP/CPAP: 0 to 25 mbar

Maximal frequency: 200 bpm

Inspiratory time:      0.1 to 2 seconds

Expiratory time:       0.2 to 30 seconds

Inspiratory flow:       1 to 30 L/min

Base flow: 1 to 30 L/min

Inspiratory Oxygen concentration: 21 -100 vol. % oxygen

Peak Pressure: to 99 mbar

Mean Airway Pressure: to 99 mbar

Monitoring:

Flow Monitoring:      Should be at Y-piece, integrated

Volume Monitoring: Should be at Y-piece, integrated

Lung function monitoring:       Compliance, resistance C20/C, time constant, r

FiO2 Monitoring:     Integrated

Real-time curves:

Flow and pressure, integrated

Graphic trends:

6 parameters, integrated

 

 

Mechanical Ventilator with Air Compressor

 

Patient type: Pediatric, Adult

Control Principle:     Must be time-cycled, volume-controlled, pressure limited

Gas Supply: Must have Turbine technology

Turbine Exchange: Exchange interval must be 8 years, with no limit in operating hours during this interval

Flow sensor Type: Must be temperature-compensated hot-wire anemometer

Ventilation Modes:

Standard modes:     VC-CMV/VC-AC, VC-SIMV, SPN-CPAP

Optional modes:      VC-MMV, PC-APRV, PC-BIPAP/PC-SIMV+, PC-AC

Optional Enhancement:

• AutoFlow: Must have Automatic adaption of the inspiratory flow in volume orientated ventilation modes.

• NIV        Must be capable in providing Non Invasive Ventilation with optimized alarm systems and automatic leakage compensation.

• Capnography: Must be using Mainstream CO2 measurement

• MonitoringPlus      Must have Loops, Trends, user Logbook

• LPO      Must have Low Pressure Oxygen. Independent oxygen supply, e.g. with an O2 concentrator

• Nurse call  Must have Connection for transmitting alarm signals to a central, alarm system

Advanced Features:

• Oxygen enrichment – provides 180 seconds of 100% O2 prior to suction procedure

– interrupts ventilation and delivers a minimal flow to detect reconnection automatically

• Manual Inspiration/Inspiration Hold Provides manual ventilation

• Day/Night screen switch-over

Selection of two (2) types of screen for day and night mode

• Key pod lock prevents unwanted modification of settings

• Standby mode Interrupts ventilation and provides access to other ventilator options and settings

 

Alarms:    Must have an Audible and visual alarms with event log and priorities

Performance data:

Trigger:

Leak adapted Trigger: Flow/volume trigger

Setting Parameters:

Peak inspiratory pressure:      10 to 80 mbar

PEEP/CPAP: 0 to 25 mbar

Respiratory Rate: 2-80bpm for Setting Value

0-50 bpm (measured value)

Inspiratory time:   0.2 to 10 seconds

Expiratory time:       0.2 to 30 seconds

Inspiratory flow:       1 to 30 L/min

Base flow: 1 to 30 L/min

Inspiratory Oxygen concentration: 21 -100 vol. % oxygen

Peak Pressure: to 99 mbar

Mean Airway Pressure: to 99 mbar

Monitoring:

Flow Sensor is located expiratory valve not in Y-piece

Y-piece flow sensor is for neonatal

Lung function monitoring:       Compliance, resistance C20/C, time constant, r

FiO2 Monitoring:     Integrated

Real-time curves:

Flow and pressure, integrated

Graphic trends:

6 parameters, integrated

 

/2018/bid bulletin/#32/p5

 

 

  Logbook:

Record of last 100 alarms

Secondary screen display:

With 3 curves displayed (full shaded curves)

With Loops display

At least large color touch screen display with mouse and keyboard

Waveform Storage:

With at least 12 hours full disclosure

With graphic trend display of ventilation up to 10 days

With waveform review for recording three real time curves up to 12 hours

Trends table display:

10 measured display and set values as well as alarms messages can be tracked over periods of 10 days

Logbooks: Stores up to 100 alarms, warnings and advisory messages

Patient data:

Can input patient name, age, gender and room ID number

Storage:  Internal hard disk at least 30G data storage

USB port for: Mouse, keyboard, printer

Special Features:

With Hardrive

USB ports

Large colored touched screen display

Second monitor must be detachable to the ventilator unit

With internal hard disk for storage

Brand must be in the local and international market for at least 10 years

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 30-60 days with demonstration

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

Logbook:

Record of last 100 alarms

*OPTIONAL: *Secondary screen display:         

     *With 3 curves displayed (full shaded curves)           

     * With Loops display      

     *At least large color touch screen display with mouse   

      and  keyboard

Waveform Storage:

With at least 12 hours full disclosure

With graphic trend display of ventilation up to 10 days

With waveform review for recording three real time curves up to 12 hours

Trends table display:

10 measured display and set values as well as alarms messages can be tracked over periods of 10 days

Logbooks: Stores up to 100 alarms, warnings and advisory messages

 

 

Storage:  Internal hard disk at least 30G data storage

USB port for: Mouse, keyboard, printer

Special Features:

With Hardrive

USB ports

 

Second monitor must be detachable to the ventilator unit

With internal hard disk for storage

Brand must be in the local and international market for at least 10 years

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 30-60 days with demonstration

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

 

8

 

Patient Monitor with ECG, Heavy duty

 

Modular Type of Monitor

Five (5) Parameters:

1. ECG 3 Leads – with Arrhythmia Analysis (Asystole, Bradycardia, Tachycardia, Ventricular fibrillation, Ventricular tachycardia)

2. SpO2 – One (1) Unit – SpO2 Sensor for Pedia/Adult water proof and shock proof

3. Non- Invasive Blood Pressure – Oscillometric with step down deflation using Dinamap Superstat Technology (dual tubing)

Note: BP Cuff Set (Neonate, Pedia and Adult)

4. Respiratory Rate

5. One (1) Temperature Port

 

Non-Invasive Blood Pressure setting has 4 Cycle Setting

Uses Dinamap Technology for the Non Invasive Blood Pressure monitoring

Four Simultaneous lead arrhythmia detection for confirmation and elimination of false alarms

Central Station Ready– Optional Upgrade

10.4” TFT LCD Display

Resolution SVGA (800 x 600)

Displays: 6 waveforms and 4 digit fields

Alarm System with Three Level Priority Prompt with patient and system status categories

Brand must be in the local and international market for at least 10 years

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 60 days

Allocation:

CPH-Balamban = 1; CPH-Bogo = 1; CPH-Danao = 1

 

NO CHANGES

 

/2018/bid bulletin/#32/p6

 

9

 

Suction Machine Table Top, heavy duty

 

with Table/Cart

1.5L or more capacity

Vacuum adjustment allows for 80-550mm with double pump brushless DC Motor

 

Delivery: To be delivered directly to allocated hospital within 60 days

 

Allocation:

CPH-Balamban = 1

CPH-Bogo = 1

CPH-Danao = 1

 

NO CHANGES

 

10

 

Syringe Pump heavy duty

 

infusion modes including multi-step mode, delayed start mode and interval / intermittent mode which allows the appropriate infusion therapy for the patient.

 

Midpress system

Midpress peristaltic function

With Anti Free Flow (AFF) function, prevents inadvertent excess infusion by automatically closing the tube with the anti-free flow clip when the door is opened.

 

can detect any brand of IV sets

Volume Limit Range: 1-9, 999ml, in 1 mL increment

LED status indicators

Has alarm: sound and light when there is any unusualities during infusion

 

Brand must be in the local and international market for at least 10 years

Warranty: 1 year

Delivery: To be delivered directly to allocated hospital within 60 days with demonstration

 

Allocation:

CPH-Balamban = 1

CPH-Bogo = 1

CPH-Danao = 1

 

NO CHANGES

 

For guidance and information of all concerned.

 

Cebu Capitol, Cebu City, May 11, 2018.

 

(sgd)      ATTY. MARK C. TOLENTINO

Chairman

Conformed:

 

______________________________________

(Signature over printed name)

 

______________________________________

(Name of Company)

 

Date: _____________________________________

 

**Please acknowledge receipt hereof by re-sending this through Telefax No. (32) 253-9614, duly signed received.**

 

 

TO DOWNLOAD THE FILE, PLEASE CLICK:
Adden. No. 18-32 - for Adv. 18-056_ICU EQPTS